Flow Rate Display; Re-Spiking An Irrigation Bag - Stryker AHTO Manual De Utilización Y Mantenimiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 51
8. Press the blue button for irrigation. Press the red button to aspirate.
Aspirate
WARNING The system is intended for short bursts of irrigation, with a

Flow Rate Display

The Flow rate display shows the current flow rate of the Irrigation Pump. When
the system is priming, the display will flash. When the cassette is inserted and the
irrigation button is not depressed, the display will remain solid.
Display
C
FC
E1
E2
E3
L
M
H

Re-Spiking an Irrigation Bag

When re-spiking a bag, perform the following steps:
1. Remove the AHTO tube set cassette from the AHTO Irrigation Pump.
2. Re-spike the bag.
3. Reinsert the cassette into the pump.
10
maximum irrigation time of 15 seconds every 2 minutes.
High temperatures may result if the system is activated for
longer periods.
Table 1: Flow Rate Display
Cassette not present; System is charging
Cassette not present; Unit is fully charged
Rotation of motor impeded
Batteries are empty
Cassette not present; Batteries not plugged in or improperly connected
Power cord is plugged into unit
2 L / min.
3 L / min.
4 L / min.
Irrigate
Meaning

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

250-070-601

Tabla de contenido