6. PARAMETER SETTING
6.1 PARAMETER SETTING
ENTER THE MENU
000
ENTER THE SELECTED MENU
SELECT THE MENU
ESCAPE FROM THE MENU
To access the technical parameters,it's required to press the calibration switch, switch is locating below the housing.
Only authorized personal. Break the seals produces device loose its verification.
Boîtier (couvercle avant)
Joint d'étanchéité (caoutchouc)
Joint d'étanchéité (acier inoxydable)
Bouton d'étalonnage à l'intérieur
Vis M4 (boîtier fixé)
Boîtier (couvercle arrière)
Turn on the scale.
Press TARE key during that start up until display will be shown P1---;
Then input password, the default password is 000;
Press G/N key to move the active digit;
Press TARE key to increment a digit;
Press ZERO key to confirm, display will shows F0 CAL, scale enter setting mode
Press ZERO key, it can confirm which will be shown displayed.
Press TARE key, it can choose menu block one by one.
Press G/N key to escape from the setting menu.
Fil de cuivre
Plomb
USERS GUIDE SMC
EN
EN