DETAILED OPERATING INSTRUCTIONS
(CONT.)
AUTO PRIMING
Automatic priming results in a small
amount of fluid being discharged
beyond the delivery set adapter.
12. Open both clamps on the delivery
set. Remove the luer adapter cap
from the delivery set, and place
the cap in the luer adapter cap
holder available on the pump.
13. Press PRIME for automatic prim-
ing of the delivery set. The Pump
displays a series of Ps during the
prime cycle and the rotor begins
to turn. Feeding is primed first,
then water (if necessary) through
the delivery line. This fluid is not
counted in Volume Delivered.
The prime cycle completes in
approximately 3 minutes. An
audible alarm sounds when prim-
ing is complete, the pump goes to
Pause mode, and the Feeding rate
flashes.
USE WITH ENTRAFLO™ PUMP
SETS
If desired, EntraFlo™ Pump adminis-
tration sets may be used to deliver
water or feeding only.
1. For standard set delivery, use only
the feeding side. This allows rate
and dose programming.
2. Install set according to instructions
on page 9.
3. Select rate. Press
change.
4. Check Volume Delivered and to
clear press reset.
or
to
BEGIN FEEDING
1. Connect delivery set adapter to
the patients enteral feeding tube.
CAUTION: Verify enteral tube
placement before beginning feeding.
2. Press RUN/PAUSE to begin
enteral feeding. The Pump pri-
marily displays the feeding rate.
It briefly displays the feeding vol-
ume delivered every 7 seconds.
3. Press RUN/PAUSE and/or OFF
to stop the pump at any time
(Feeding rate, Dose Limit, Feeding
and Water Volume Delivered,
Feeding and Water Accumulated
Volume, Set Volume, and Time
Interval are retained in pump
memory).
Pressing SELECT allows the user to
scroll through and view Feeding and
Water side settings at any time.
5. Select Dose Limit. Press
change.
6. Select Accumulated Volume and to
reset press reset.
7. Set Strength at FS.
8. Deactivate Hydration and
Programmable Flush by select-
ing Time Interval and press until
nOnE is displayed.
12
or
to