CONTROLS
FEEDING SIDE DISPLAYS
24. FEEDING RATE, displays the
feeding infusion rate. Select from
1–295 mL/hr in 1 mL increments.
25. VOLUME DELIVERED, displays
full strength or diluted feeding
volume delivered.
26. DOSE LIMIT, displays dose limit.
Select from 1–2000 mL in 1 mL
increments. Dose is completed
when Volume Delivered equals
the Dose Limit set. Refer to
Dose Delivered under Alarms/
Indicators. If nOnE is displayed,
this feature is deactivated.
27. ACCUMULATED VOLUME, dis-
plays full strength or diluted
feeding volume delivered over
multiple feedings.
28. STRENGTH, (Feeding Dilution)
displays feeding strength to be
delivered. Select from FS (Full
/
Strength),
3
,
4
strength, or H
(CONT.)
/
/
/
/
2
,
1
,
1
,
1
3
2
3
4
0 (water only).
2
WATER SIDE DISPLAYS
29. SET VOLUME, displays the vol-
ume of water to be delivered
over the selected Time Interval
or at the end of the dose limit.
Select from 0–500 mL in 1 mL
increments.
30. TIME INTERVAL, displays the
time interval (2 to 12 hours in 1
hour increments) over which the
selected water volume is deliv-
ered. When doSE is displayed,
indicates Programmable Flush
is activated. When nOnE is dis-
played, all water side settings
are deactivated.
31. VOLUME DELIVERED, displays
amount of selected water vol-
ume that has been delivered.
At the conclusion of the Time
Interval selected, the Volume
Delivered turns to 0 and a new
accumulation begins.
32. ACCUMULATED VOLUME, dis-
plays water volume delivered
over multiple Time Intervals.
8