QUICK REFERENCE

13
(HOW TO SET FEEDING RATE, DOSE LIMIT,
OR RESET FEEDING VOLUME DELIVERED)
Note: Changes can only be made to parameters at the START of a new feeding
protocol to avoid improper water delivery. To facilitate this safety feature,
parameters can only be changed when amount displayed in Volume Delivered
is 0.
If desired, reset Volume Delivered or reset all parameters to default settings
by pressing and holding Reset for 4 seconds when Rate is displayed.
TO SET THE FEEDING RATE
1. Press SELECT until Feeding Rate
is displayed.
2. Enter a new feeding rate by
pressing
(decrease). Change is more rapid
when the key is pressed and
held.
3. Adjustable in 1 mL/hr increments
from 1–295.
TO SET THE DOSE LIMIT
1. Press SELECT until Dose Limit is
displayed.
2. Enter a new dose limit by press-
ing
rapid when the key is pressed
and held.
3. Adjustable in 1 mL increments
from 1–2000.
4. For continuous feeding when no
dose limit is desired, press
until nOnE is displayed.
(increase) or
or
. Change is more
TO RESET VOLUME
DELIVERED AND/OR
ACCUMULATED VOLUME
If Feeding Strength is activated
to deliver diluted feeding, Feeding
Volume Delivered and Accumulated
Volume Delivered display the TOTAL
DILUTED FEEDING (feeding AND
water) delivered.
1. Press SELECT until Volume
Delivered is displayed.
2. Press RESET.
3. Follow steps 1 and 2, using
SELECT until Accumulated
Volume is displayed, and reset if
desired.
Any time Feeding Volume Delivered
is reset, Water Volume Delivered
is automatically reset. Likewise, if
Water Volume Delivered is reset,
Feeding Volume Delivered is auto-
matically reset.
Accumulated Volume displays full
strength or diluted feeding volume
delivered over multiple dose limit
feedings.