SELF TEST

The major functions of the pump may be
tested by the following steps:
Turn the pump on by pressing ON.
1.
.
.
.
.
2.
8
8
8
8 is displayed immediately.
While displayed, press and simulta-
neously hold the
4 seconds. The display changes to
either a 1, 2, or 3 (the dose delivered
alarm mode).
3.
Press and simultaneously hold the
and
keys again for 4 seconds.
4. After 4 seconds, an audible alarm will
sound, "teSt" is displayed momen-
tarily, and then all eights (

FACTORY SERVICE

All service work should be performed
by authorized personnel only. If the
pump does not function properly:
Contact your supplier of the
EntraFlo H20™ Nutrition Delivery
System to arrange for Pump return
and replacement.
If your pump was supplied directly
from Medline, contact your Medline
Sales Representative or call
1-800-Medline from anywhere in the
United States or 1-800-396-6996 if
calling from Canada.
If it becomes necessary to return a

SAFETY AND WARNINGS

NOT FOR PARENTERAL USE
INTENDED EXCLUSIVELY FOR
ENTERAL TUBE FEEDING AND PATIENT
HYDRATION.
Do not use this device near flammable
anesthetics, it is a possible explosion
hazard.
This device is designed to minimize the
effects of uncontrolled electromagnetic
and
keys for
.
.
.
.
8
8
8
8).
5.
The pump rotor should begin to turn.
To test the visual alarms, press either
SELECT, RESET,
RUN/PAUSE. When any of these
keys are pressed or the sensor is
momentarily blocked, all visual alarm
indicators should light, an audible
alarm sounds, and the rotor will halt
momentarily.
6. Any failure indicates a service check
must be performed. See the Factory
Service section for instructions on
where to send the pump.
7.
Turn the pump off by pressing OFF.
pump for factory service, a replace-
ment pump will be sent to you.
Carefully pack the defective pump
in the carton the replacement pump
arrived in and ship according to
instructions given by the Medline
Sales Representative.
Unauthorized repairs void warranty
coverage (see section on limited war-
ranty).
interference and other types of interfer-
ence from external sources. Avoid use of
other equipment that may cause erratic
operation or degradation in the perfor-
mance of this device.
If the integrity of the external protective
conductor in the installation is in doubt,
equipment shall be operated from its
internal electrical power source.
, PRIME, or
,