INSTALLING THE DENTAL UNIT
E0
BASE
Net weight ...... 68
Gross weight .. 90
3
.................0,6 m
cm. 125 x 69 x 70
E1
E2
INSTRUMENT TABLE
Net weight ...... 22
Gross weight .. 33
............. 0,428 m
3
cm. . 88 x 76 x 64
The dental unit is supplied into two packing cases: the
first one labelled "BASE" contains the body, the assistant
arm and the accessories; the second one labelled
"INSTRUMENT TABLE" contains the operator retraction
arm, the modules and the tray holder.
Provide air and water hose couplings with taps. It is
absolutely necessary to clean out air and water pipes
before connecting the dental unit to them.
In addition to the guard of the connecting box, we
recommend that you remove even the lid of the chair
parallelogram, in order to operate easily when introducing
hoses and cables into the dental unit. Remove such
mentioned lid by unscrewing the two long brass nuts
which are placed under it and which fasten it to the upper
rods of the parallelogram.
The chair shall be fastened to floor by inserting expansion
bolts into the two holes made on the front side of the base
footboard.
Now, mount the body of the dental unit on the chair
support by means of the stud bolts and the socket head
screws supplied (placed inside the "ACCESSORY" box,
into the packing case labelled "BASE").
To make such operation easier, fit the additional stud
bolt supplied, with the other stud bolt into the lower left
hole.
Level the body properly (horizontally and vertically)
before tightening the screws and the middle nut; remove
the Allen-head stud bolt and replace it with the third
screw supplied.
Mount the protection guard by means of the two supplied
screws.
34
ABSOLUTE
Installation instructions