Gconnecting The Suction Pump; Harrangement Of Calbenium Kit - eurodent ABSOLUTE Guia De Instalacion

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

G
CONNECTING THE SUCTION PUMP
G1
6 7
6 7
H
ARRANGEMENT OF CALBENIUM KIT
If the Calbenium mixing tank or the surgical kit is fitted afterwards, the electronic system should be set for recognizing
the unit.
Proceed as follows:
introduce the "A" jumper on the body p.c.b. with off device;
H1
switch on the dental unit keeping the clock key
if the Calbenium mixing tank is assembled, the parameter increment key
appears;
if the surgical kit is assembled,
When picking up any of the two suction tubings, the
dental unit provides a contact between the two white
wires (i.e. a switch which conducts when a tubing is up
and cuts off when tubings are at rest).
Therefore, the pump must be provided with control
switchboard.
In case of pumps as Dürr or Cattani equipped with
control switchboard, which require a simple switch
between two preset contacts (generally 1 and 3), connect
such contacts to the two brown ones 6 and 7 in front of
the white wires, as the figure shows.
pressed;
should be set. Press enter key
41
ABSOLUTE
Installation instructions
appears on the screen;
should be pressed until
to store the setting.

Publicidad

Tabla de contenido
loading

Tabla de contenido