eurodent ABSOLUTE Guia De Instalacion página 36

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

E6
E7
E8
Fasten the tray holder under the module-holder table (to
make it easier, use a 17 box or socket spanner).
Now, insert the cable and the pin of the assistant arm into
the proper housing (at the bottom of the body, on the
chair side), and lock the pin with the snap ring and its lock
space washer. In case of METASYS suction, the amalgam
separator bowl should be removed
Pass the cable in the body until it is coupled into the
connector X13 of the body p.c.b.
NOTE: ensure that the cable does not hang under the
body!
If the dental unit is equipped with assistant syringe,
follow the steps below:
Insert the syringe hose into its housing on the tubings carrier,
and make it run upwards through the hole between the
tubing couplings. Then, lock the end bush with the clip
supplied.
Cut the (green) water hose of the operator's table, fit the
T-coupling between the two pieces and connect to the
syringe water hose. The air hose has to be connected to
the coupling with plug in the hydropneumatic unit.
If the Calbenium mixing or spray feeding unit is installed, cut
the pressurization air hose, interpose the T-coupling between
the two pieces and connect to the syringe air hose.
36
ABSOLUTE
Installation instructions
temporarily.

Publicidad

Tabla de contenido
loading

Tabla de contenido