1 Warning and safety information
Electromagnetic compatibility (EMC)
Allocation of acquisition system to patient
Hygiene information
Disturbance of electronic devices worn on the
patient's body
Interference due to magnetic field of Sensor Safe
6
The SIDEXIS – Intraoral complies with the requirements
of IEC 60601-1-2:2001.
Medical electrical devices are subject to special precau-
tionary measures regarding EMC. They must be
installed and operated as specified in the document
"Installation Requirements".
Portable and mobile HF communication devices can
influence medical electrical equipment. The use of
mobile telephones in the practice or hospital area there-
fore must be prohibited.
Please also observe the ESD protective measures in
Chapter 1.2.
Within the scope of practice operations, unambiguous
allocation of the acquisition system to the examinee
must be ensured to guarantee reliable allocation of X-ray
exposures to the patient data saved by SIDEXIS!
The protective covers must be exchanged for each new
patient and the sterilizable accessories must be steril-
ized to prevent any transmission of infective agents
which might cause serious illnesses.
Suitable hygienic measures must be taken to prevent
cross contamination among patients, users and other
persons.
To prevent the malfunctioning of electronic devices and
data memories, e.g. radio-controlled watches, telephone
cards etc., these objects must be removed prior to the X-
ray exposure.
Cards with a magnetic stripe, e.g. credit cards or trans-
portation tickets, may be damaged by the magnet in the
Sensor Safe.
Avoid, for example, carrying objects that are sensitive to
magnetic fields and the Sensor Safe in the same bag or
pocket.