Sirona Dental Systems GmbH
8
Help Messages H3
H3
06
13
14
15
60 82 023 D3297
D3297.201.01.11.02
You want to release an exposure but the Ready-LED on
Multitimer is still blinking:
CAUTION
In case of unit failure the test button T on the Multitimer
must be pressed first (radiation protection measure!).
•
Press the X-RAY exposure button on Multitimer.
H3 help message appears on the kV/mA display.
•
Read on the following list what is to be done to get
the unit ready for the exposure.
•
Before carrying out the corresponding indication,
press return button R on the Multitimer to acknowl-
edge the help message.
Help Messages H3
H3 01 Press one of the R buttons to return the rotation
unit to the start position (Help-LED 16).
H3 02 No cassette inserted (Help-LED 15).
H3 03 Exposed film inserted (Help-LED 15).
H3 04 Cassette holder not in Panorama position
(Help-LED 13).
H3 05 Set the diaphragm on the wheel corresponding to
the exposure program (Help-LED 18).
H3 06 Engage the locking button on diaphragm wheel
properly (Help-LED 18).
H3 07 Move forehead support to permitted position
(Help-LED 14).
H3 20 Press the R button on the Multitimer to confirm
the exposure data (Help-LED 17).
16
17
18
19
8 Help Messages H3
41