Sirona Dental Systems GmbH
E... error messages
E3 22
Light localizer key on control panel A was actuated dur-
ing a start-up or before selftest of the unit finished.
E3 23
R key on control panel A was actuated during a start-up
or before selftest of the unit finished.
E3 27
Forehead support prevented from moving away
E3 28
column, towards
E3 29
The software endswitch was reached before the zero
position.
One of the forehead support adjustment keys ← or → on
E3 31
control panel A was actuated during a start-up or before
selftest of the unit finished.
E3 32
X-ray head was prevented from leaving panorama posi-
tion at start of exposure.
E3 33
X-ray head prevented from returning to zero position
after exposure.
E3 36
Panoramic cassette holder swings out of panorama
position during exposure.
E3 47
Program card incorrectly inserted.
60 82 023 D3297
D3297.201.01.11.02
column (possibly stuck).
Remedy
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display.
from
Remove obstruction.
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display.
Remove obstruction.
Press R key on Multitimer to reset display.
Remove obstruction.
Press R key on Multitimer to reset display.
Swing panoramic cassette holder into panorama posi-
tion.
Remove exposed film from the cassette and insert a
fresh film.
Slide loaded film cassette into the cassette tray until it
engages.
Press R key on Multitimer to reset display.
Check that program card is correctly inserted.
i
NOTE
In case of an error message not contained in this list,
switch the unit OFF and ON again.
CAUTION
Should the error occur repeatedly, call the service tech-
nician, quoting the error message.
11 List of Error Messages