B) Recharging; Troubleshooting - CefarCompex Mi-Theta 600 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

B) Recharging

Remove all stimulation cables from the
stimulator before recharging it. Connect
the charger to the mains and then con-
nect the stimulator to the charger. The
charge menu shown below will automati-
cally appear on the screen.
The duration of the recharging operation
is displayed on the screen. The charger
supplied with the stimulator fully
recharges the stimulator within 2-2,5
hours.
The battery icon is animated while
the battery is recharging. When fully
charged, the icon will be full and the
total time taken to recharge the battery
will flash on the screen. Simply discon-
nect the charger and the Mi Theta 600/
Theta 500 will turn off automatically.

6 - Troubleshooting

Electrode error
The Mi Theta 600/Theta 500 will beep
and alternately display a pair of elec-
trodes and an arrow pointing to the
channel where the problem was detect-
ed. In the example above, the stimulator
has detected an error in channel 2.
The electrode error message can mean
that:
• there are no electrodes connected to
this channel,
• the electrodes are old, worn, and/
or the contact is poor: try using new
electrodes,
• the electrode cable is faulty: try con-
necting it to another channel. If the
problem persists, replace the cable.
If for any reason yor device seems to
present a malfunction other than those
presented above, please contact Cefar-
Compex customer service.