Maintenance:
Gross Particle Intake Filter
Fine Particle Intake Filter
Patient tubing
Humidifier chamber

CLEANING

WARNING
To avoid electric shock, always unplug the power cord from the wall outlet power source before cleaning.
CAUTION–Never rinse or place the device in water. Never allow liquids to get into or around any of the ports, switches or
air filter. Doing so will result in product damage. If this occurs, do not use the device. Contact your equipment provider for
service.
Device Cabinet
1.
Unplug the SleepCube from the power source. Wipe the cabinet with a clean, damp cloth every few days to keep the
cabinet dust free.
2.
Allow the device to dry completely before plugging it into a power source.
Filters
Air-Inlet Filter
1.
The air-inlet filter should be checked every 10 days and cleaned as necessary. The filter is located in the rear of the
device.
CAUTION–Proper filter function is important for the operation of the device and to protect the device from damage.
2.
To clean, remove the dark outer foam filter from the frame.
3.
W ash the filter in a solution of warm water and dishwashing detergent and rinse with water. Allow the filter to dry
completely. If the filter (part #DV51D-602) is damaged, contact your equipment provider for a replacement.
4.
Ensure the filter is completely dry before installation and use.
Fine Particle Filter (Optional)
1.
If you have an optional fine particle filter (part #DV51D-603), inspect this filter and replace it if it shows signs of dirt or
damage. Do not wash.
2.
Install the fine particle filter behind the standard filter.
NOTE–If the fine particle filter is not installed first according to the directions, the life of the filter will be reduced requiring
more frequent replacement.
Optional Heated Humidifier
Water Chamber ‑ Humidifier water chamber should be cleaned daily.
1.
Press on the chamber release latch and pull out the water chamber from the cradle (Fig 4).
2.
Separate the chamber halves by pressing forward on the chamber disassembly lever and lifting the lid of the
chamber (Fig 9).
3.
Remove the sealing gasket from the chamber lid (Fig 10).
4.
Hand wash the parts in a solution of warm water and mild dishwashing detergent.
CAUTION–Do not use any solutions that contain bleach, chlorine, alcohol, scented oils or other aromatic-based ingredients
to clean the humidifier components or tubing. These solutions may reduce the life of the product.
5.
Allow the parts to dry before reassembly.
CAUTION–Inspect the gasket seal before each use to ensure that it is properly seated in the groove around the entire
chamber.
NOTE– The disassembled chamber may also be washed once per week in a dishwasher on the delicate or glassware
cycle, top shelf only.
Humidifier Cradle ‑ The humidifier cradle should be cleaned as needed.
1.
Turn off the heater and unplug from power source. Allow system to cool for at least 10 minutes.
EN - 12
Inspection Interval
10 days, clean if dirty
10 days, replace if dirty
Daily
1 / month
Replacement Interval
6 months
30 days
6 months
1 year
Specification
> 3.0 micron particles
> 0.3 micron particles
See tubing spec
SE-DV54-1