REFERENCE CONDITIONS
Temperature
Relative humidity
Supply voltage
Auxiliary resistances R
, R
H
AC and DC interference voltages
Electric field
Magnetic field
METROLOGICAL SPECIFICATIONS
Measurement range
(automatic range)
0.00...19.99 Ω
20.00...199.9 Ω
200.0...1999 Ω
Response time
DISTORTION QUANTITIES
Temperature
Relative humidity
Operating voltage
Voltage circuit load (1)
Current circuit load (2)
DC voltage
AC interference voltages on H
AC interference voltages on S
(1) Resistance measured or resistance of rod E + resistance of rod S + resistance of leads
(2) Resistance measured or resistance of rod E + resistance of rod H + resistance of leads
(3) Possible flashing of HIGH RESISTANCE indicator above 4.5 V
4. SPECIFICATIONS
and R
S
E
Resolution
Measurement current
(square signal 128 Hz)
0.01 Ω
10 mA
0.1 Ω
1 mA
1 Ω
100 µA
between 4 and 8 s depending on measurement conditions
Operating limits
-10...+55°C
8.8...12.5 V
0.00...19.99 Ω
20.00...199.9 Ω
200.0...1999 Ω
0...20 V (3)
0...23 V
0...9 V
20
23°C ± 3 K
45% < RH < 55%
10.5 V ± 0.2 V
nil
nil
< 1 V/m
< 40 A/m
Accuracy
± 2%rdg ±1 pt
± 2%rdg ±1 pt
± 2%rdg ±3 pt
Max variation
± (1% rdg ± 1 pt) per 10°C
20...90%
± 2% rdg ± 5 pt
± (1% rdg ± 1 pt) per Volt
50 kΩ
± 1% rdg per 10 kΩ ± 4 pt
30 kΩ
± 1% rdg per 10 kΩ ± 2 pt
30 kΩ
± 1% rdg per 10 kΩ ± 2 pt
50 kΩ
± 1% rdg per 10 kΩ ± 2 pt
negligeable
or 32 V
± 2% rdg ± 2 pt
rms
peak
or 13 V
± 0.5% rdg ± 2 pt
rms
peak
Empty voltage
≤ 42 V
peak
≤ 42 V
peak
≤ 42 V
peak