The main components are as follows (figure 1):
Jaws:
Clamping capacity:
Current direction:
Sizes:
Output:
gap: 112 mm (1). Capacity with jaws open: 250 mm.
cable with a maximum diameter of 115 mm (2).
arrow (3), visible on the side of the clamp, indicating the current's direction.
The current is deemed to be circulating in the positive direction when it
flows from the current's generator to the current's consumer.
2
- Measurement current: 4 A: output 1 mV/mA.
- Measurement current: 400 A: output 1 mV/A.
voltage output, via a 1.5 m-long cable moulded into the clamp and
terminating in two 90° elbow male safety plugs.
16