Temperature Measurements; Mv Measurements; Troubleshooting; Specifications - YSI EcoSense pH100 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

3. If AUTOLOCK is off, the pH value of the sample displays on the screen. If both pH and
temperature readings are stable, take a reading. If AUTOLOCK is on, press MEA./EFF.
"WAIT" flashes until the unit determines a stable pH reading.

Temperature Measurements

The model pH100 can measure temperature independently with the ATC/Temp probe without
using the pH electrode. Place the ATC/Temp probe in the media to be measured. The measured
temperature displays.

mV Measurements

1.
Connect the optional combination mV electrode to the unit. Press MODE to enter mV mode
with AUTOLOCK on or off as desired. For inherently unstable samples, the unit will not
AUTOLOCK. Use mV mode with AUTOLOCK off in this case.
2.
Rinse electrode with distilled water and immerse it in sample to be measured. If AUTOLOCK is
off, the mV value of the sample will be displayed on the screen. If AUTOLOCK is on, press
MEA./EFF. "WAIT" flashes until the unit determines a stable mV reading.

TROUBLESHOOTING

MAIN
DISPLAY
• pH electrode offset is greater/less than
Er 1
• STAND was pressed before the
• pH electrode is faulty
• pH electrode slope is off by more than
Er 2
• SLOPE was pressed before the
• Buffer 4.00, 4.01, 9.18 or 10.01 is not
• Temperature is out of the 0.0 to 60.0 °C
Er 3
• Measured pH is out of the 16.00/-2.00
OvEr/Undr
• Measured mV is out of the 1250/-2000
• Measured temperature is out of the

SPECIFICATIONS

Display
pH
mV
Temperature °C
POSSIBLE CAUSE
+/-1.5 pH.
electrode and ATC/Temp probe settle to
within ±1.5 pH of the buffer value.
30% of ideal slope.
electrode and ATC/Temp probe settled
to within 30% of the buffer value.
correct.
range.
pH range.
mV range.
10/120 °C range.
Range
-2.00 to 16.00 pH
-1999 to 1250 mV
-10.0 to 120 °C
• Replace the buffer and/or the
pH electrode. Press ESC.
• Allow sufficient time for the
electrode and ATC/Temp probe
to stabilize.
• Return for service.
• Check that the correct buffer is
used and that the electrode
slope is not off by more than
30% from the theoretical slope.
• Allow sufficient time for the
electrode and ATC/Temp probe
to stabilize.
• Replace the buffer and/or the
pH electrode. Press ESC.
• Return for service.
• Bring the buffer temperature
within range.
• Return for service.
• Bring the out of range unit into
the correct measuring range.
• If units are within proper range,
return product for service.
-
Accuracy
±0.1%, ±2 lsd
±0.1%, ±1 lsd
±0.5 °C
6
CORRECTIVE ACTION
Resolution
0.01 pH
1 mV
0.1 °C

Publicidad

Tabla de contenido
loading

Tabla de contenido