3
Clean Transmitter
a. Prepare
• Protect: Wear clean gloves and goggles
• Prepare Soak: Put Clorox Healthcare
deep enough to submerge the transmitter
b. Clean
• Rinse and Scrub: Rinse transmitter in cold tap water while
brushing with a soft bristle brush until all visible soil is gone
• Soak and Scrub:
• Put transmitter in prepared soak for 3 minutes
• While immersed, scrub uneven areas (see green arrows)
with a soft bristled brush or a pre-saturated bleach wipe
c. Rinse and Dry
• Rinse: Remove from soak and rinse transmitter under flowing cold tap water for 10 seconds
• Dry: Wipe transmitter dry with cloth
d. Inspect
• Verify there is no visible soil. If there is, clean again.
Bleach Germicidal Cleaner solution (Clorox) in a container
®
3