4
Disinfect Transmitter
a. Prepare
• Protect: Wear clean gloves and goggles
• Prepare Soak and syringe:
• Put CaviCide
• Fill syringe with about 30 mL of Cavicide
b. Disinfect
• Flush:
• Focus on the uneven areas
• Swirl in Cavicide for 10 seconds
• Refill syringe
• Scrub:
• Saturate clean cloth or wipe with Cavicide
• Wipe entire transmitter for at least 3 minutes or until all soil is removed
• Focus on the uneven areas
• Flush:
• Focus on the uneven areas
• Swirl in Cavicide for 10 seconds
• Soak:
• Put transmitter in prepared soak
• Swirl it for 30 seconds
• Then let it soak for another 3 minutes
c. Rinse and Dry
• Rinse: Remove from soak and rinse transmitter under flowing cold tap water for 10 seconds
• Dry: Wipe transmitter dry with cloth
4
solution (Cavicide) in a container deep enough to submerge the transmitter
®