Low Pressure
Indicator
See "Troubleshooting and Alarm Conditions" on page 15 for
possible causes of Low Pressure.
Service Indicator/Pump Fault
Power Fail Indicator
Weight settings are affected due to power fail. There are two
variants of pump in the market as follows:
1. The pump will revert to a default patient weight setting of
125 kg (275 pounds) once power is restored after a power
outage or generator test.
2. The pump will revert to the last setting set by the caregiver
once power is restored after a power outage or generator test.
To determine which pump is being used once the pump is
operating, set the pump to 20 kg (44 pounds) and disconnect
from the power supply. Reconnect power and if the pump
reverts to 125 kg (275 pounds), you have pump 1, and if it
reverts to 20 kg (44 pounds), you have pump 2.
8