9. Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing,
or immersing the sensors in liquid may cause inaccurate readings.
10. SpO
measurements may be adversely affected in the presence of high ambient
2
light. Shield the sensor area if necessary.
11. Excessive patient movement may cause inaccurate readings.
12. Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter,
or intravascular line may cause inaccurate readings.
13. Fingernail polish or false fingernails may cause inaccurate SpO
Contraindication
14. Significant levels of dysfunctional hemoglobins (such as carbonxy-hemoglobin or
methemoglobin) may cause inaccurate readings
15. Intravascular dyes such as indocyanine green or methylene blue may cause
inaccurate readings
16. Venous pulsations may cause inaccurate readings.
17. Hypotension, severe vasoconstriction, severe anemia or hypothermia may cause
inaccurate readings.
18. Cardiac arrest or shock may cause inaccurate readings.
No side effects known
Follow local ordinances and recycling instructions regarding disposal or
recycling of the device and device components, including batteries.