•
The Oil Test Set must be properly earthed.
•
The test chamber must be kept scrupulously clean.
•
If the test chamber cover is cracked or damaged in any way, the test set
must not be used, but sent for repair to an authorised agent.
•
A routine continuity check between the emc/discharge barrier and the earth
point should be made weekly (see page 9, section 5).
•
The correct type of oil test vessel must be used, and correctly positioned on
the support horns before carrying out any testing (see page 9, section 7).
•
Replacement fuses must be of the correct type and rating.
•
Safety warnings and precautions must be read and understood before the
instrument is used. They must be observed during use.
THE TEST SET MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS
Symbols used on the instrument
2
SAFETY WARNINGS
Caution: Refer to accompanying notes.
Risk of electric shock.
Equipment complies with relevant EU Directives
Earth terminal
NOTE
.