Commissioning; Checking The Installation; Connecting The Power Supply - ABB Sensyflow FMT500-IG Instrucciónes De Puesta En Marcha

Ocultar thumbs Ver también para Sensyflow FMT500-IG:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 159

Commissioning

6
Commissioning
The device may only be started up / opened by qualified operating personnel. The device must
be installed and the electrical signal lines must be connected prior to start-up.
6.1

Checking the installation

Prior to start-up, check that the equipment has been installed correctly:
Is the device securely fastened?
Have all the electrical signal, control and interface cables been laid and connected correctly?
6.2

Connecting the power supply

Connecting the 115 / 230 V power supply
Step
1.
2.
3.
4.
Connecting the 24 V power supply
Step
1.
2.
3.
Warning
When connecting the power supply, the information provided below must be observed.
Failure to observe the information provided can result in severe injuries or death.
Important (Notice)
When using a device with an explosion-proof design, comply with the explosion protection
regulations. For Sensyflow FMT500-IG values, please refer to the relevant certificates (ATEX,
FM, CSA, GOST).
Before connecting the power supply, check that the set operating voltage matches the line
voltage.
Before making any other connections, you must first establish a connection between the
protective-conductor terminal and a protective conductor.
A line switch with adequate switching capacity and capable of fully isolating the device from
the supply system (all poles) must be installed within reach of the installation location. The
protective benefits of the protective conductor must not be canceled out as a result of this.
The protective conductor must not be interrupted or detached either inside or outside the
device.
34 - EN
Action
Check whether the voltage specified on the name plate matches the line voltage.
Use a power supply line with sufficient dimensions and ratings (circuit breaker).
In the vicinity of the device, install a supply circuit isolator in the power supply
line or a switched socket so that the device can be fully isolated (all poles) from
the power supply. Label the supply circuit isolator to make it clear that it relates
to the equipment that needs to be isolated.
Connect the power cord to the power supply.
Action
Check whether the voltage specified on the name plate matches the line voltage.
Use a power supply line with sufficient dimensions and ratings (circuit breaker).
Connect the supply line to the power supply.
Sensyflow FMT500-IG
CI/FMT500-IG-X1

Publicidad

Tabla de contenido
loading

Tabla de contenido