ResMed AirStart10 APAP Guia Del Usuario página 16

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

 
Environmental conditions
Operating temperature:
Operating humidity:
Operating altitude:
Storage and transport temperature:
Storage and transport humidity:
Electromagnetic compatibility
The AirStart 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-
2:2007, for residential, commercial and light industry environments. It is recommended that mobile communication devices
are kept at least 1 m away from the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on
www.resmed.com, on the Products page under Service and Support.
IEC 60601-1:2005 classification
Class II (double insulation), Type BF, Ingress protection IP22.
Sensors
Pressure sensor:
Flow sensor:
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
O for more than 6 sec or 40 cm H
2
Sound
Pressure level measured according to ISO 17510-1:2007 (CPAP mode):
SlimLine:
Standard:
Power level measured according to ISO 17510-1:2007 (CPAP mode):
SlimLine:
Standard:
Declared dual-number noise emission values in accordance with ISO 4871:1996.
Physical - device and water tub
Dimensions (H x W x D):
Air outlet (complies with ISO 5356-1:2004):
Weight (device and standard water tub):
Weight (device and cleanable water tub):
Housing construction:
Water capacity:
Standard water tub - material:
Cleanable water tub - material:
Temperature
Maximum heater plate:
Cut-out:
Maximum gas temperature:
14
+41°F to +95°F (+5°C to +35°C)
Note: The air flow for breathing produced by this therapy
device can be higher than the temperature of the room.
Under extreme ambient temperature conditions
(104°F/40°C) the device remains safe.
10 to 95% relative humidity, non-condensing
Sea level to 8,500' (2,591 m); air pressure range 1013 hPa to
738 hPa
-4°F to +140°F (-20°C to +60°C)
5 to 95% relative humidity, non-condensing
Internally located at device outlet, analog gauge pressure
type, -5 to +45 cm H
Internally located at device inlet, digital mass flow type, -70
to +180 L/min
O for more than 1 sec.
2
26.6 dBA with uncertainty of 2 dBA
26.6 dBA with uncertainty of 2 dBA
34.6 dBA with uncertainty of 2 dBA
34.6 dBA with uncertainty of 2 dBA
4.57" x 10.04" x 5.91"
(116 mm x 255 mm x 150 mm)
22 mm
40.5 oz (1147 g)
40.5 oz (1147 g)
Flame retardant engineering thermoplastic
To maximum fill line 380 mL
Injection molded plastic, stainless steel and silicone seal
Injection molded plastic, stainless steel and silicone seal
154°F (68°C)
165°F (74°C)
 106°F ( 41°C)
O
2

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Airstart10 cpap

Tabla de contenido