Initiating X-Ray Exposure - VATECH PaX-i3D Smart Manual Del Usuario

Ocultar thumbs Ver también para PaX-i3D Smart:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 236
3.3
PaX-i3D Smart(PHT-30LFO) User Manual
Completing Patient Positioning
1. After confirming patient positioning and beam alignment, adjust the temple
supports to fit snugly on both sides of the patient's head by clicking the temple
support adjustment button.
Make sure that the Temple Support is in Close position
before clicking the READY button.
CAUTION
2. Click the Ready button after the patient has been properly positioned. No
X-Ray will be emitted at this point.
3. Proceed to section 3.3 Initiating X-Ray Exposure.

Initiating X-Ray Exposure

The X-Ray exposure method and its sequences are the same for PANO
Standard and PANO Special modes. The example used in this manual is the
PANO Standard mode.
To emit the X-Ray, perform the following procedure.
-
Do not operate the PC during exposure. Failure to
comply with this instruction may cause the system to
malfunction.
-
The operator shall observe the X-Ray safety regulations
WARNING
applicable to their area at all times during the operation
of this equipment.
If an emergency occurs during image acquisition, release
the exposure switch to cease X-Ray emission.
CAUTION
Direct the patient to not move during the exposure until the
Temple support is open.
IMPORTANT
3 Acquiring PANO Images
35

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pht-30lfo

Tabla de contenido