3 Acquiring PANO Images
Exposure indicator
light turns orange
Orange: Exposure
X-Ray ON
36
1. Leave the X-Ray room and close the door. The operator must visually
monitor the patient at all times during image acquisition.
2. Press and hold down the exposure switch until image acquisition is complete.
3. Release the exposure switch when "Image capturing is completed" message
appears on the screen.
4. The image appears in real time on the imaging GUI.
5. Click the Save button to save the image.
After Image Acquisition
After acquiring the image, complete the following steps:
■ Loosen the temple supports to release the patient.
■ Remove the hygiene cover from the bite block (for standard PANO mode
only).
■ Press Return to bring the Rotating Unit back to its initial position.
During X-Ray irradiation, ensure that
- The LED lamp on top of the equipment
t u r n s o r a n g e t o i n d i c a t e X - R a y
emission.
- The warning lamp outside the X-Ray
room turns on.
- The sound (beep or music: optional)
goes off.
- The radiation symbol on the upper left
corner of GUI turns yellow to indicate X-
Rays are being emitted.