5.3
Exposure indicator
light turns orange
Orange: Exposure
X-Ray ON
PaX-i3D Smart(PHT-30LFO) User Manual
Initiating X-Ray Exposure
-
Do not operate the PC during exposure. Failure to
comply with this instruction may cause the system to
malfunction.
-
The operator shall observe the X-Ray safety regulations
WARNING
applicable to their area at all times during the operation
of this equipment.
Direct the patient to not move during the exposure until the
Temple support is open.
IMPORTANT
1. Leave the X-Ray room and close the door. The operator must visually
monitor the patient at all times during image acquisition.
2. Press and hold down the exposure switch until image acquisition is complete.
3. Release the exposure switch when "Image capturing is completed" message
appears on the screen.
4. The image appears in real time on the imaging GUI.
If an emergency occurs during image acquisition, release
the exposure switch to cease X-Ray emission.
CAUTION
5 Acquiring CT Images
During X-Ray irradiation, ensure that
- The LED lamp on top of the equipment
turns orange to indicate X-Ray emission.
- The warning lamp outside the X-Ray
room turns on.
- The sound (beep or music: optional) goes off.
- The radiation symbol on the upper left
corner of GUI turns yellow to indicate X-
Rays are being emitted.
59