DISPLAY AND KEYBOARD
O
Zero Indicator
: Shows that
the weighing plate is empty.
Before weighing a product,
be sure that the weighing
plate is empty and press
"O" key for zero setting if
the O indicator is off.
Otherwise error may occur.
Equilibrium Indicator D : Shows that
the load on the weighing
plate is well stabled. It is off
while placing or removing the
product. If
indicator
D
goes off, the weight shown
on
the
display
may
be
wrong. Therefore, in order to
read the weight, wait until
symbol appears after
D
placing the product on the
weighing plate.
NET
NET Indicator
: It is ON, when there
is a tare. Actual weight value
is NET weight value.
T
Tare Indicator
: Shows that
the tare effects the weighing.
It flashes when the tare is
not fixed. It is continuously
on when the tare is fixed.
(Only for scale with LCD
display.)
M
Total Memory Indicator
: It flashes
when the total memory is
containing
a
value.
remains
continuously
on
while the content of the total
memory is being displayed.
R
Battery Indicator:
: Shows that
the battery is used.
Battery indicator is ON, scale
is powered by the battery. If
it flashes, it is necessary to
charge or change the battery
in the shortest time.
Weighing
instrument
designed with using LCD
type display modules has
four
additional
indicators for indicating the
fullness of the battery and
also they have a sliding
battery charging animation
mode,
which
activated when the adapter
connection is established.
While the adapter connection
is
not
established,
battery frame around the
battery indicators appears for
indicating the battery is being
used. Just before the battery
becomes empty, the battery
frame flashes.
W1
W2
and
Indicator:
Shows actual
weighing range (only for
multiple range scales). See
max, min, e values of each
range on data plates.
Key : Used for accepting the
T
load as tare.
Key : Used for zero setting
O
the scale when the weighing
plate is empty.
Key : Used for reaching
F
setting
menu
and
functions similar to settings.
Key : Used for adding total
It
W
memory.
M / X
Key : Used for displaying
and deleting total memory.
☼
Key : Toggles backlight
If
between ON, OFF, ON-OFF
(only for LCD displays)
HOLD MODE: This feature valid for
unapproved devices. Hold mode is
- 9 -
entered by pressing and holding the
backlight
appears on the screen.In this mode,
that
pressing the backlight key shortly,
Weight, unit price, total displays are
fixed for seconds, and Weight screen
battery
blinks .By pressing and holding the
backlight button to exit from this mode
and "HLdOff" appears on the screen.
is
only
FIRST OPERATION
When scale is turn on all segments and
symbols
are
backwards are seen on display about 10
the
seconds. After that, maximum capacity and
division is seen for a while and scale gets
ready to weigh.
If there is any weight on weighing plate or
there is no weighing plate on scale, ER 5
error message might be seen on display. Be
sure that there is no weight on the weighing
plate before calling service.
WEIGHING THE PRODUCT
Place the product on the weighing plate.
some
Wait until the
the product is seen on display.
The value "0.000" should appear on the
display and also the
should be on when the weighing plate is
empty. If the value on the display is not
zero(or
indicator is not on), press O
D
for zero setting. After establishing zero and
balance, proceed to the weighing process.
WARNING: Absolutely not place any load
more than maximum capacity of scale.
button
and
"HLdOn"
("Max" value on data plate)
Product should be placed on weighing plate
by hand. Leaving the product from high or
throwing it on to the weighing plate (even the
weight is low) scale may be damaged.
o
seen.
Numbers
counting
o
Value of zero is seen
on display. O and
indicators are
D
ON. Nothing should
be
placed
on
weighing plate until it
o
gets
ready
for
weighing operation.
o
NOTE: Press T key 2 times for blocking
the cancellation of tare. In this case, word
indicator is on. Weight of
D
sabit
"
while and tare symbol (only for scale with
LCD display) begins to be continuously ON.
To cancel tare, press T key when the
and O indicator
D
weighing plate is empty.
Models with T symbol (for LCD display);
• Never ON
• FLASHING
- 10 -
WEIGHING WITH TARE
Place the tare weight on the
weighing plate. Wait until the
D
indicator is on.
Weight of tare can be read
from weight display.
Press T key.
"dara" is seen on display
for a second then value of
0.000
"
"
is
seen
on
display.. NET indicator is ON.
T
symbol is flashing (only
for LCD display). Tare has
been weighted.
Place the product on weighing
plate. Wait until
indicator is
D
on.
Product's net weight is seen
on display screen.
Make empty the weighing plate.
Wait until the
indicator is
D
on. NET and
T
indicators
off.
Tare
is
cancelled
automatically.
" is seen on weight display for a
: There is no tare
:When weighing plate is
empty,
tare
will
be
cancelled
automatically.