NOTE:
• Do not press the punctured site against the test strip or try to smear the blood.
• If you do not bring the strip to the blood sample within 3 minutes, the meter
will automatically turn off. You must remove and reinsert the test strip to start
a new test.
• The confirmation window should be filled with blood before the meter begins
to count down. NEVER try to add more blood to the test strip after the drop of
blood has moved away. Discard the used test strip and retest with a new
one.
• If you have trouble filling the confirmation window, please contact your health
care professional or the local customer service for assistance.
26
4. Bring strip to sample
While holding the meter in an angled
position with the strip facing down,
gently bring the strip to the sample so
that the absorbent hole can sip in the
blood. Confirmation window should be
completely filled if enough blood sample
has been applied. Do NOT remove your
finger until you hear a beep sound.