Important Control Solution Information - Fora TN'G Voice Manual Del Propietário

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

● Do not bend, cut, or alter a test strip in any way.
● Keep the strip vial away from children since the cap and the test strip
may be a choking hazard. If swallowed, promptly see a doctor for help.
For further information, please refer to the test strip package insert.

Important Control Solution Information

● Use only our control solutions with your meter.
● Do not use the control solution beyond the expiration date or 3 months
after first opening. Write the opening date on the control solution vial
and discard the remaining solution after 3 months.
● It is recommended that the control solution test be done at room
temperature 68°F to 77°F (20°C to 25°C). Make sure your control
solution, meter, and test strips are at this specified temperature range
before testing.
● Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
● Store the control solution tightly closed at temperatures between
35.6°F to 86°F (2°C to 30°C). Do NOT freeze.
36

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido