► Micro-Kill+™ (Micro-Kill Plus™) by Medline
(EPA Reg. No. 59894-10-37549)
To obtain disinfecting wipes and other information, please contact
Medline at 1-800-MEDLINE (1-800-633-5463) or visit www.medline.com.
Disinfecting Procedures
1. Take out one disinfecting wipe from the package and squeeze out any
excess liquid in order to prevent damage to the meter.
3. Remove the wipe. Allow the meter surface to dry completely.
4. Discard the used wipes and never reuse them.
Wash your hands thoroughly with soap and water after handling the
blood glucose monitoring system, lancing device and test strips to avoid
contamination.
2. Wipe all exterior surface of the meter
including the display and buttons.
Hold the meter with the test strip slot
pointing down and wipe the area around
the test slot, be careful not to allow
excess liquid to get inside. Keep the
meter moist with disinfection solution
contained in the wipe for a minimum of
2 minutes for Micro-Kill+™ wipes. Follow
the instructions on the package label of
disinfecting wipe. Use two or more wipes
if necessary.
33