IV. 11. Troubleshooting
Observations
The 0-I button is in the ON
position. The device does not
operate. The continuous alarm
sounds.
 
 
Oxygen concentration indicator
remains lit yellow.
The alarm test does not work.
The 0-I button is ON and the
compressor is operating but
there  is  no  flow.  The  audible    
alarm sounds.
The 0-I button is ON, the
compressor is operating, there
is  a  flow  but  the  audible  alarm  
sounds.
The compressor stops in mid-
cycle, then starts again after a
few minutes.
The  oxygen-­enriched  air  flow  is  
interrupted at the nasal
cannula outlet.
The  flow  at  the  nasal  cannula    
outlet is irregular.
AirSep
Corporation
®
Probable causes
Power cable not plugged in
correctly. Mains power failure.
 
 
Oxygen concentration is too
low.
Internal electrical fault.
Pneumatic connection broken
or compressor problem.
 
Internal electrical fault.
Pneumatic circuit fault.
 
Compressor thermal safety
device has been tripped. Fan
not working.
Tube  disconnected  or  humidifier  
not tight.
Pneumatic  circuit  problem.    
 
 
 
 
 
 
Solutions
Check the cable connection.
Reset the circuit breaker (2) if
necessary by pressing. Check
the fuses or circuit breaker
fitted  on  the  premises.
Contact your distributor.
Contact your distributor.
Stop the device by pressing the
0-I button and contact your
distributor.
Stop the device and contact
your distributor.
Stop the device and wait for it
to cool down. Check that the
patient circuit is not obstructed.
Start up again. If the device
does not start, contact your
distributor.
Check  the  gas  administration
circuit.
Contact  your  distributor.
GB
F
9-GB