Coltene CanalPro CL Instrucciones Para El Uso página 21

Ocultar thumbs Ver también para CanalPro CL:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 38
(3) Mounting and Removing File
Insert file to contra angle head, lightly turn the file until it engages with the latch mechanism. Push it in-
ward to click. File Removal, depress the push-key and pull out the file.
Fig. 7
CAUTION
- When attaching and detaching the file, turn off the power beforehand.
- After the file is locked in place, lightly pull out the file to make sure the file is locked.
- Always clean the shank of the file to be installed before use. Allowing dirt to enter the chuck could cause
loss of concentricity and deterioration of chucking force.
- Do not exceed the rotation speed recommended by the file manufacturers.
(4) Preparatory Operations
1. Hold down the POWER key for more than one second to
turn on the power.
2. Keep holding down the PROGRAM key until program
No. suitable for file to be used is displayed.
3. When changing the set value of speed, torque, gear ra-
tio or auto reverse, press the SELECT key, select the set-
ting item and press the UP/DOWN key to change.
NOTICE
The GEAR RATIO key is fast-forwarded, when it is continu-
ously held down.
(5) Operation
If you press the ON/OFF key briefly, the motor handpiece starts. If you re-press the key, it stops. (Alternate
operation)
If you hold down the ON/OFF key for more than one second, the motor handpiece starts while the key is
pressed. If you release the key, it stops. (Intermittent operation)
40
Fig. 8
<Operation Panel>
ON/OFF key
Liquid crystal panel
SELECT key
PROGRAM key
UP key
DOWN key
POWER key
„ Auto Reverse Function
Auto reverse setting can be changed by pressing the SELECT key for one second or longer while the motor
handpiece stops. As the auto reverse mark flashes while the SELECT key is kept pressed, press the UP/
DOWN key to adjust it.
AUTO REVERSE
AUTO STOP
(flashes while the SELECT key is kept pressed, and goes out when it is
Auto Reverse OFF
released)
The alarm will sound when it becomes a value of half of a torque limit set up during motor handpiece rota-
tion, and the alarm will change near the torque limit value. (Load alarm) When it continues applying a load
and it exceeds the torque limit value you can select mode from the following three modes.
AUTO REVERSING
The handpiece stop that it reaches for the loaded value, the file rotates in reverse. When a load is removed,
the file returns to normal rotation (forward rotation) automatically
Load lower than
Load higher than
the set torque limit value
the set torque limit value
Forward rotation
Reverse rotation
AUTO STOP
The motor handpiece starts in reverse. When a load is removed, the motor handpiece stops. If you want it
to rotate (forward-rotate) again, re-press the ON/OFF Key or re-step.
Load lower than
Load higher than
the set torque limit value
the set torque limit value
Forward rotation
Reverse rotation
Reverse rotation when load continues,
forward rotation when load is removed
Returns to forward
Fig. 9
Reverse rotation
when load continues,
stop when load is removed
Stop
Fig. 10
41

Publicidad

Tabla de contenido
loading

Tabla de contenido