4.2 Multiple patient use
Cleaning and disinfection cycles
Nebuliser with
PARI SINUS tube system
Compressor housing
Air filter
Nebuliser
The nebuliser is not suitable for use by more than one patient.
Use a separate nebuliser (including connection tubing) for each
patient.
Disinfecting the compressor housing
For disinfection, use a standard, alcohol-based disinfectant
(e.g., isopropanol). Regarding the use and quantity of the disin-
fectant, it is imperative to follow the information for use of the
agent used.
Disinfect the compressor as follows:
• If the compressor is visibly dirty, clean the compressor be-
fore disinfecting it [see: Cleaning the compressor housing,
page 28].
• Moisten a cloth with the disinfectant.
NOTE! Danger of device fault due to liquid penetration.
Never spray any liquids onto the compressor or the power
cord. If liquid does get into the compressor, it must not be
used under any circumstances. Before starting the com-
pressor again, contact the manufacturer or distributor.
• Wipe the outer surfaces of the housing off thoroughly with
the cloth.
– 29 –
Replace before each patient change
Disinfect before each patient change
Replace after 200 operating hours
(approx. 1 year)