7
With the tester still attached, fully submerge the transmitter in the enzymatic
solution for one minute.
8
Holding the tester, remove the transmitter from the solution. Brush the entire
surface of the transmitter using a soft-bristled brush, paying close attention to
hard-to-clean areas, until visibly clean.
9
Rinse the transmitter under room temperature tap water until all visible detergent
is gone.
-15-