Applied parts
Operating conditions
Storage conditions
Battery life
Transmitter frequency
Transmitter size
Transmitter weight
Charger size
Charger weight
Transmitter
Sensor
Transmitter Temperature: +32° to +122°F (0° to 50°C)
Caution: When operating the transmitter on a tester in air
temperatures greater than 106°F (41°C), the temperature of
the transmitter may exceed 109°F (43°C).
Transmitter Relative Humidity: 10% to 95% with no condensa-
tion
Charger Temperature: +50° to +104°F (10° to 40°C)
Charger Relative Humidity: 30% to 75% with no condensation
Transmitter Temperature: -4° to +131°F (-20° to +55°C)
Transmitter Relative Humidity: 10% to 100% with condensation
Charger Temperature: +14° to +122°F (-10° to +50°C)
Charger Relative Humidity: 10% to 95% with no condensation
Transmitter: 14 days of continuous glucose monitoring immedi-
ately following a full charge
Charger: Uses one new AAA to charge the transmitter
MMT-7703NA 916.5 megahertz
MMT-7703WW 868.35 megahertz
The dimensions of the transmitter are approximately:
1.13 x 1.41 x 0.38 inches
The weight of the transmitter is approximately 5.4 grams
The dimensions of the charger are approximately:
1.75 x 2.5 x 1.15 inches
The weight of the charger is approximately 25 grams
-21-