Immunity Test
IEC 60601 Test
Radiated RF
3 V/m
IEC
80MHz to
61000-4-3
2.5GHz
Compliance
Level
Level
3 V/m
Electromagnetic Environment -
Guidance
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the MiniLink
transmitter, including cables, than
the recommended separation dis-
tance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation dis-
tance:
d=1.2√P 80MHz to 800 MHz
d=2.3√P 800MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmit-
ter manufacturer and d is the recom-
mended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
be less than the compliance level in
each frequency range. Interference
may occur in the vicinity of equip-
ment marked with the following sym-
bol:
-24-
a
, should