Intraoral equipment suggested exposure times,
in seconds and impulses
DC SUPPLY
AC SUPPLY
Lower incisor/cuspid
Lower premolar
Lower molar
Upper incisor/cuspid
Upper premolar
Upper molar
Reference conditions:
• Adult patient
• Anodic current 7 mA.
• Source to Detector Distance SSD = 250 mm (10 inches)
• Total (inherent) filtration 2 mm Al equip.
Unless the image is grossly over- or under-exposed, image density can be corrected by sub-
sequent software processing (contrast/brightness — see Imaging Software User Manual and
Installation Guide).
1
One impulse is equal to .017 sec. at 60 Hz. The recommended number of impulses can be
adjusted as necessary for proper exposure, resulting in adequate image density and contrast
and low noise gratuity.
Panoramic/cephalometric equipment
The DenOptix QST system requires no adjustments to the panoramic/cephalometric exposure
settings (kV, mA or time). The DenOptix QST system can be used with almost any existing
panoramic/cephalometric unit, regardless of age or manufacturer.
The use of a cassette without an intensifying screen is highly recommended. If you choose
to use a cassette with an intensifying screen, you must at a minimum remove the front
screen. If the intensifying screens are not removed, you will get a low quality image due
to the shielding effect from a decreased dose reaching the imaging plate. To remove the
intensifying screens, follow the directions below.
User and Service Manual
50 kV
60 kV
0.32 sec.
0.25 sec.
19 impulses
15 impulses
0.40 sec.
0.32 sec.
24 imp.
19 imp.
0.50 sec.
0.40 sec.
30 imp.
24 imp.
0.40 sec.
0.32 sec.
24 imp.
19 imp.
0.50 sec.
0.40 sec.
30 imp.
24 imp.
0.64 sec.
0.50 sec.
38 imp.
30 imp.
28
Gendex DenOptix® QST
60kV
65 kV
65 kV
70 kV
0.16 sec.
0.16 sec.
1
20 imp.
10 imp.
0.16 sec.
0.16 sec.
25 imp.
10 imp.
0.20 sec.
0.20 sec.
32 imp.
12 imp.
0.16 sec.
0.16 sec.
10 imp.
10 imp.
0.25 sec.
0.20 sec.
15 imp.
12 imp.
0.32 sec.
0.25 sec.
19 imp.
15 imp.
Doc # M010-004WWE July 2005
70 kV
75 kV
0.12 sec.
7 imp.
0.12 sec.
7 imp.
0.16 sec.
10 imp.
0.12 sec.
7 imp.
0.16 sec.
10 imp.
0.20 sec.
12 imp.