Cleaning the carousel
Figure 7-1
Remove the intraoral imaging plate holder from
the carousel by removing the three screws
along the vertical plastic retention piece. After
removing the three screws, the intraoral plate
holder will come free from the carousel.
Figure 7-2
Spray the soiled surfaces of the imaging plate
holder and carousel with a general household
non-abrasive cleaner or soap and water. Do not
use strong alkaline or ammonia based cleaners.
Wipe clean and allow to air dry.
Figure 7-3
Insert the dry imaging plate holder under the
vertical plastic retention piece on the carousel.
Slightly tighten the screws to hold the first side
in place. Take the other side of the intraoral
plate holder and place under the plastic reten-
tion piece and tighten the screws.
42
Doc # M010-004WWE July 2005
User and Service Manual