Maintenance procedures

The DenOptix QST System is designed for many years of trouble-free operation. It is manufac-
tured from the highest quality components to ensure excellent performance. Maintenance
that can be performed by the operator is minimal.
7.1

Cleaning the system

If the scanner, imaging plates or carousel becomes contaminated, clean in accordance with the
directions given in Section 5.6. If, however, a component merely becomes soiled, clean as follows:
Cleaning the scanner
Turn off the DenOptix QST scanner, as described in Section 5.7, before cleaning. Wipe the
outside surfaces with a paper towel dampened with a cold sterilant solution or household,
non-abrasive cleaner (window cleaner works well). DO NOT SPRAY OR SOAK THE SCANNER.
Wipe the inside of the carousel well with a very slightly damp cloth. Be careful not to allow
running or dripping solvents into the DenOptix QST scanner. This could cause damage to the
electronics inside. Allow to air dry before plugging in or turning back on.
Cleaning imaging plates
Imaging plates should be handled carefully. For the best images, take care not to scratch them
and keep them dust-free. Use the following procedure to clean them:
1. Use lint-free, 100% cotton gauze (not cotton balls). Gently wipe the cotton gauze over the
dry imaging plate surface. Wipe back and forth and then in a circular motion.
2. To clean any remaining stains, dampen the gauze in anhydrous (water-free) isopropyl
alcohol and wipe using the same motion as above.
3. Completely dry the surface by wiping with another piece of cotton gauze. Ensure that the
imaging plate is completely dry before use.
User and Service Manual
41
7
Doc # M010-004WWE July 2005