15. Do not put hands onto the rotor unless the rotor is completely
stopped.
16. Do not move unit while it is operating.
17. Do not immerse the product in water. Injury due to shock or fire may
result.
18. The centrifuge rotors must be loaded symmetrically. Operating
this unit with an unbalanced rotor will cause severe vibration and
damage, which is not covered by warranty.
19. Do not attempt to use centrifuging tubes with the magnetic stirrer
rotor.
20. Do not fill tubes while inserted in centrifuging rotors. Spilled liquid
may harm unit.
21. Close tube lids before starting centrifuge. Open lids can be torn off
during centrifugation and damage the unit. Open lids can cause fluid
to be dispersed inside the centrifuge.
22. Do not insert non-standard centrifuge tubes or other foreign articles
into the rotor.
23. Follow all precautions from the Material Safety Data Sheet (MSDS)
for any reagent you use with this instrument.
24. Do not use solvent or flammable liquids near this or other electrical
equipment.
25. Do not centrifuge or stir flammable, explosive, or corrosive materials.
26. Do not operate in a hazardous or flammable environment.
27. Do not use any solvent on the unit that may attack plastic or cause
cracks in the rotors or corrode metal parts in the rotors.
28. Ensure rotors are protected from corrosion and mechanical damage
at all times. Rotors must be cleaned with a pH-neutral cleaning liquid
and immediately dried.
29. Always work in a manner which endangers neither the user nor any
other person.
30. Should the instrument fail to work to specification, immediately
stop using it. Clean and troubleshoot the instrument according to
the instructions under "Trouble Shooting" before any further use of
the instrument. Contact an authorized dealer or the manufacturer if
trouble persists.
31. Do not attempt to stop a rotor while unit is running. Doing so may
cause the unit to fail and will void the warranty.
32. Only use original manufacturer's rotors, AC adapters, tube adapters
and any other spare parts.
33. Repairs are to be performed by trained and authorized service
personnel only. Contact Heathrow Scientific
34. Opening the instrument housing or improper use of the instrument
voids the warranty. If there is a failure during the warranty period,
contact Heathrow Scientific
35. Do not attempt to use units that have not been correctly installed or
repaired.
4
LLC for warranty service.
®
LLC.
®