7.3
Microscope assembling
7.3.1 B-510POL
1. Insert the Bertrand lens ① on the frame and
lock the locking screw ② with the provided Allen
wrench. (Fig. 1)
2. Insert the optical head on the Bertrand lens and
lock the locking screw with the provided Allen
wrench. (Fig. 2)
•
Hold the head with one hand during the locking
in order to avoid that the head falls.
3. Insert both eyepieces into the tubes of the optical
head. (Fig. 3)
•
One of the two eyepieces is equipped with
a crosshair for centering the entire optical
system. It is advisable to insert the eyepiece
with crosshair in the right eyepiece sleeve.
•
Condenser is pre-installed in the factory. To
remove the condenser use an Allen wrench
1,5mm diam and operate on the locking screw
placed on the right side of the condenser holder.
4. Screw each objective into the thread of
the
nosepiece,
magnification. (Fig. 4)
clockwise
with
increasing
①
Page 8
②
F ig. 1
F
ig. 1
F ig. 2
F
ig. 2
F F ig. 3
ig. 3
F ig. 4