How to clean and disinfect the meter?
The meter must be cleaned prior to the disinfection. Use one
disinfecting wipe to clean exposed surfaces of the meter thoroughly and
remove any visible dirt or blood or any other body fluid with the wipe.
Use a second wipe to disinfect the meter. Do NOT use organic solvents
to clean the meter.
We recommend for meter cleaning and disinfection you should use the
disinfecting wipe/towelette from below. The following product has been
shown to be safe for use with the FORA GTel Multi-Functional
Monitoring System.
► Micro-Kill+™ (Micro-Kill Plus™) by Medline
(EPA Reg. No. 59894-10-37549)
To obtain disinfecting wipes and other information, please contact
Medline at 1-800-MEDLINE (1-800-633-5463) or visit
www.medline.com.

Disinfecting Procedures

1. Take out one disinfecting wipe from the package and wipe all meter's
exterior surface display and buttons.
2. Keep the meter surface wet with disinfection solution for a minimum
of 2 minutes for Micro-Kill+™ wipes. Follow the instructions on the
package label of disinfecting wipe.