Use each test strip immediately after removing it from the vial.
Write the opening date on the vial label when you first opened it.
Do not use test strips beyond the expiration date. This may cause
inaccurate results.
Do not bend, cut, or alter a test strip in any way.
Keep the strip vial away from children since the cap and the test strip
may be a choking hazard. If swallowed, promptly see a doctor for
help.
For further information, please refer to the test strip package insert.
Important Control Solution Information
Use only FORA GTel control solutions with your meter.
Do not use the control solution beyond the expiration date or 3
months after first opening. Write the opening date on the control
solution vial and discard the remaining solution after 3 months.
It is recommended that the control solution test be done at room
temperature 68°F to 77°F (20°C to 25°C). Make sure your control
solution, meter, and test strips are at this specified temperature
range before testing.
Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
Store the control solution tightly closed at temperatures between
35.6°F to 86°F (2°C to 30°C). Do NOT freeze.
34