General Remarks - MSA AUER SSR 30/100 Manual Del Usuario

Autorescador de oxígeno
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 19
inhalt 06/03
13.06.2003 11:02 Uhr
– Breathing resistance
(inhalation and exhalation)
– Oxygen content:
– Carbon dioxide contents:
– Temperature of inhalation air: (30 to 55) °C
– Service time under escape
conditions
(Air consumption 35 l/min):
*) The individual service time can differ, depending
on the air consumption. For example, at an air
consumption of only 10 l/min (e. g. while
seated with donned apparatus while waiting for
rescue) the service time can reach up to 100 min.
6
Use
Opening the container (Fig. 2)
Push locking device towards arrow until lead seal
breaks and lock falls off.
Pulling out the SSR (Fig. 3)
Grasp container cover with one hand and with the
other hand pull out the rescuer by the white neck
strap. If, when opening the container, the cover
falls away with the locking device, the rescuer will
automatically unfold and be suspended by the
mouthpiece plug. Pull apparatus from mouthpiece
plug.
Fitting neck strap (Fig. 4)
If a protective helmet is being worn, take it off prior
to placing neck strap around neck, replacing it
after neck strap is in place.
Insert mouthpiece (Fig. 5)
Place mouthpiece into mouth with flange of mouth-
piece between teeth and lips. Bite down on the
two lugs. Inflate breathing bag with exhalation air.
Fit nose clip (Fig. 6)
Grasp nose clip by rubber nipples, pull nose pads
apart and fit them unto nose.
Adjust neck strap and tie waist belt (Fig. 7)
Adjust neck strap, ensuring that the head is able to
move freely. Fit waist belt and tighten with knot.
7
Inspection, Maintenance and
Reconditioning
7.1

General Remarks

To assure proper functioning the MSA AUER SSR
30/100 must be regularly checked for exterior
damage and cleaned if necessary (do not use a
metal brush!). The apparatus requires no further
maintenance work.
Since it contains a humidity sensitive chemical,
the apparatus is packed airtight in a container and
lead sealed at the factory. The apparatus check is
made on the closed container. The check intervals
depend on how the apparatus have been treated,
in accordance with the intervals schedule (7.2).
Seite 5
On used apparatus and apparatus that have been
(3 to 7) mbar
opened the chemical canister must always be
replaced. A reuse is not permitted for safety
> 21 vol. %
reasons.
approx. 1 vol. %
Interior checks and reconditioning of the
ready-for-use device shall be performed by
MSA AUER or authorized service centers.
The following rules must be observed for the
20 min *)
return of used equipment for overhauling:
• The apparatus may only be packed cooled off
and dry.
• Organic substances (e. g. solvents, alcohols,
oils, greases, fuels, etc.) must not get into the
chemical canister.
If the complete unit cannot be packed in the
container, the following procedure is required:
• The chemical canister must be removed from the
apparatus (loosen clamp, remove protective
cover), packed inside the container and sealed
with the original locking assembly of the
container (do not use finely cut filling material).
• The other apparatus components are placed
with the closed container and all are carefully
packed.
• The components must be complete and
undamaged.
Replacement of components increases the repair
cost.
7.1.1 Disposal
Apparatus no longer required must be disposed of
properly.
The chemical canister contains strong alkaline and
oxidizing substances (potassium hydroxide,
potassium carbonate and potassium peroxide).
For disposal local regulations are to be observed.