10. Technical Data
Device class 93/42/EEC
Dimensions (WxHxD) in mm
Weight (without water)
Temperature range
– Operation
– Storage
Humidity during operation and storage
Ambient pressure range
Power supply*
Power consumption*
Classification according to EN 60601-1*
– Degree of protection against electric shock
Electromagnetic compatibility according to EN 60601-1-2*
– Radio interference suppression
– Radio interference resistance
Max. perm. filling level
Max. perm. operating pressure
Max. perm. flow (freely flowing)
Max. mask temperature
Gas leak at 30 hPa
Pressure drop at a flow of:
50 l/min
100 l/min
150 l/min
Resistance (inspiratory and expiratory). The use of the humidifier
has no influence on the quality of the therapy.
Flow rate: 30 l/min
60 l/min
90 l/min
User settings which result in 100% rel. humidity at the patient
connection opening at ambient rel. humidity: 70%, flow: 20 l/
min, room temperature:
15° C
18° C
23° C
VENTIclick
II a
120 x 120 x 155
265 g
+5 °C to +35 °C
–40 °C to +70 °C
≤ 95 % relative humidity
600 to 1100 hPa
40 V DC
24 VA
Type BF
EN 55011
EN 61000-3-2, EN 61000-3-3,
EN 61000-4-2 to 6
EN 61000-4-8, EN 61000-4-11
300 ml
40 hPA
300 l/min
38 °C
not measurable
0,5 hPa
1,7 hPa
3,8 hPa
0,1 hPa
0,3 hPa
0,7 hPa
setting
6
6
6
Technical Data
EN
21