S19-881

Maintenance Instructions

WARNING: Water preservatives protect potable water from bacteria, fungi and algae growth
for up to four months. If a water preservative is not used with this eye wash, it
is recommended that the tank be cleaned weekly or as often as needed.
Infection, loss of vision or other serious injury may result from misuse. DO
NOT use expired, cloudy, colored or contaminated solution. DO NOT open
preservative bottle if safety seal is broken. DO NOT save unused solution.
Step 1: Inspect eye wash unit
1. Visually inspect the eye wash unit weekly for cleanliness and proper solution level (per ANSI
Standard Z358.1-1998).
2. Activate the eye wash unit monthly to ensure proper operation. If the unit is not working properly,
remove it from the hazard area and contact Bradley Corporation immediately.
• Check the sprayheads' water spray pattern for an even, continuous flow. The sprayheads are
attached to the arm with screws and can be detached from the tray for cleaning or obstruction
removal, if necessary, and reattached (use a 9/64" hex-head Allen wrench to remove and reinstall
the sprayheads).
• Check the water supply hose at the basin of the eye wash unit for cracks, leaks or wear.
• Check the entire unit for possible water leakage.
• Check the solution for contamination and replace with fresh solution, if necessary.
Step 2: Clean eye wash unit
1. Carefully remove the eye wash tank from the pedestal and drain the water solution from the tank
(dispose of the solution properly). For ease of draining, the pedestal has a drain hole which will
accept a 1/2" I.D. tube at the underside of the hole (tube supplied by others) (see Figure 2b on
page 5 for drain hole location).
2. Clean the inside of the tank using warm water and a mild soap product; rinse thoroughly.
3. Refill the eye wash tank following the procedures outlined in Step 2 found on page 4.
• To order additional bottles of potable water preservative, contact your Bradley representative
or local distributor and ask for Bradley On-Site® Refill Kit S19-899 (kit includes bottle of
preservative, cap and liners).
4. Reinstall the eye wash tank following the procedure outlined in Step 2 found on page 4.
To operate eye wash unit
1. Begin operation by pulling down the tray. OPTIONAL UNIT
WITH HEATER JACKET: pull the front flap down and all
the way out. This action should automatically activate the
water spray (Figure 4).
2. Hold eyelids open with thumbs and forefingers. Rotate eyes
back and forth to ensure a complete rinse. Allow water to
bathe both eyes simultaneously for a minimum of 15 minutes.
3. Seek medical attention immediately after use.
CAUTION:
Do not lean on the eye wash tray when in the
opened position. Damage to unit may result.