5 Starting the DMA
12
In the measuring mode the following soft key functions are available:
Menu
Selects menus for settings and configurations. Access to a
menu can be restricted by a password.
Print
Starts a printout.
Sample#
For entering a sample text and/or number.
Method
Selects a measuring method (see Chapter 6.5).
Start
•
Starts an automatic measuring procedure. Depending on
the settings, the automatic measuring procedure includes
waiting for a stable measurement, printing the results and
storing them in the memory, freezing the display and
increasing the sample number.
•
Starts the temperature scan, if activated (DMA 5000 only).
5 Starting the DMA
Hints:
•
The DMA does not require any special installation conditions. The
installation conditions should correspond to conditions in a typical
laboratory.
•
However, to guarantee temperature stability, do not place the DMA
- near a heater
- near an air conditioner
- in direct sunlight.
Important:
A powerful built-in cooling fan dissipates heat from the base and the rear of
the DMA. Ensure that the airflow is not blocked.
Preparing the DMA for the first start-up
1. Take 2# Luer injection adapters from the accessory box.
XDLIB28