Hints:
•
•
14. As soon as the attemperating of the DMA to 20 °C is finished, perform a
Important:
The DMA is factory adjusted and this control measurement should be
performed to check if the adjustment is still valid after transport.
6 General Settings
6.1
Hint:
For a graphic overview, see the menu tree in Appendix F.
Using the Keys on the Keypad
•
•
•
•
•
XDLIB28
After turning on the power, the DMA needs approx. 20 minutes for
attemperating and additionally 5 to 10 minutes for internal temperature
adjustments. During this time "attemperating" is displayed. If the desired
measuring temperature is already set, do not touch any key during this
time as this will considerably increase the waiting period.
In case of high air humidity or low measuring temperatures see Appendix
A.
density check measurement using the supplied liquid density standards (see
Chapter 7.2).
Menu Operation
To select the main menu press the "Menu" soft key.
To select menu items use the "UP" or "DOWN" keys and press "↵" to confirm
the selection.
In the submenus found under "method settings", "output selection" and
"memory configuration" and "printer configuration" toggle between "Y" (yes,
selected) and "N" (no, not selected) using the "↵" key. Move to the next item
using the "UP" and "DOWN" keys.
In the submenus found under "method settings", "display configuration"
toggle between "N" (no, not selected), "S" (small size), "M" (medium size)
and "L" (large size) using the "↵" key.
In the other submenus, to select a menu item
-
press "↵" to activate the item,
-
move the cursor to the desired position using the "Left" or "Right" soft keys,
E N
6 General Settings
15