•
•
4. Ensure that there are no gas bubbles in the measuring cell.
5. Print your measuring results.
•
•
6. Fill in the next sample or clean and dry the measuring cell in-between, if
Hints:
•
•
XDLIB28
The filling of the measuring cell can be observed through the viewing
window. Take care that the entire measuring cell is filled with sample. A
sample amount of approx. 1.5 mL is necessary.
Leave the syringe in the filling position in order to prevent sample leakage.
Manual printout:
A manual printout is possible at any time:
•
Press soft key "Sample#" to define a sample name and/or sample number.
•
Wait until a measuring result is displayed.
•
Print the measuring results by pressing the soft key "Print".
Automatic printout and data storage:
•
If "mode" is set to "measurement valid" (in "method settings", "printer
mode") and "memory mode" is set to "measurement valid" (in "method
settings", "memory configuration") press the soft key "Sample#" and enter
a sample name and/or sample number. Measuring results are printed and
stored automatically as soon as they are valid.
•
If "mode" is set to "meas. valid (after start)" and "memory mode" is set to
"meas. valid (after start)" then press soft key "Start". The sample number
assigned via "Sample#" is displayed. An acoustic signal and the flashing
display indicate that the measurement results are valid. The measuring
results are automatically stored and/or printed.
•
Press soft key "Esc" to interrupt the measurement or soft key "Cont." to
start the next measurement. The numerical part of the sample number is
increased automatically or the next sample number from the list is selected.
necessary.
Clean and dry the measuring cell after each measurement if samples with
different chemical compositions are to be measured. For similar samples
(similar chemical composition and similar density) replace the previous
sample by rinsing the measuring cell with a sufficient amount (10 mL or
more) of the new sample.
Do not leave samples in the measuring cell longer than absolutely
necessary. Clean and dry the measuring cell as soon as possible.
E N
8 Measurements
27